Subject: G130xA-008 Supersedes: none. Product Number G1304A, G1305A, G1307A, G1308A and G1309A (rev. A.02.02) Headline An update is available for the HPLC ChemStation software which fixes two calibration related problems in revision A.02.02 of the HPLC ChemStation software. Parts required: see text below Situation: The following two problems with calibrations were observed: STARS SR# 1653-070573: If the Sequence Recalibration settings are set to: RetTime : Average or Replace Response : No Update a new level with id. "-1" is created in the calib. table instead of just updating the RetTime. STARS SR# B400-100412: Quadratic calibration point weighting does not work. (Weight set always to 1 == Equal) Solution/Action: The software was modified to make these functions work properly. This was done by producing a new revision of the quantification library and the executable file. The HPLC ChemStation software now correctly handles retention time updating and quadratic calibration point weighting. The new revision is available as an update disk and can be obtained from Product Support in WAD and LFS. The updated software reports a revision code of A.02.03. In the case of customers who observe these problems in previous revisions, they must first be upgraded to revision A.02.02 before the update is applied. Installation: To install the update, copy the file UPDATE.EXE from the diskette to the ChemStation executable directory and run UPDATE.EXE from the DOS-prompt. For HPLC 2D ChemStation software G1304A or G1305A, the executable directory is HPCHEM\LCEXE, while for HPLC 3D ChemStation software G1307A, G1308A or G1309A the directory is HPCHEM\LC3DEXE. The Update will extract new versions of HPLC.EXE and HPQUA100.DLL. When the update has been performed, the HPLC ChemStation should report it's revision as A.02.03. DO NOT attempt to install this update on revision A.01.00 or A.02.00 software! As this update is an interim revision, it has not been subjected to the same level of software testing as a full sofware release. Therefore, the division will only update customers who are extremely inconvenienced by this problem. All customers who receive the update should be upgraded to the next release when it becomes available. Date: 29-October-93 ----------------------------------------------------------------- MODIFICATION RECOMMENDED -------------------------- Standards: Labour Hours __N/A_ Travel Charge___N/A__ Action Category: On Specified Failure XX Immediately __ Agreeable Time __ Location Category: Customer Installable_XX On-Site XX HP location __ HP Responsible Until: Next release available Author/Entity: Leif Thomson/Wolfgang Winter, B400 Additional Info:__________________________________________ *****************************************************************